Measuring Range |
-50oC
-1300oC; -50oF -1999oF |
|
Sensor |
K Type
thermocouple |
|
Input |
Single(1300), Dual (1302,1303) |
|
Resolution |
-50
+ 199.9...0.1
-50
+ 199.9...0.1
(Others-1/1) |
|
Accuracy:
|
-50oC
to 199.9oC...0.3% rdg + 1oC
-50oF to 199.9oF...0.3% rdg + 2oF
-50oC to 1000oC...0.5% rdg + 1oC
-50oF to 1999oF. ..0.5% rdg + 2oF
1000oC to 1300oC..0.75% rdg + 1oC
(apply to mainframe only at 23oC +/- 5oC) |
|
Function |
H Data
Hold & Display oC/oF selection
0.1o/1oresolution selection
T1 & T2 dual points display
T1 - T2 differential measurement
3-1/2 digit LCD |
|
Temperature Coefficient |
Less
than 0.1 time the applicable
accuracy specification per
()
|
|
Overange |
1------ |
|
Low Battery |
BT |
|
Temperature Coefficient |
Less
than 0.1 time the applicable
accuracy specification per oC (oF) |
|
Input Protection |
60V DC
or 24 Vrms AC maximum
input voltage on any combination of
input Connector.
Sampling Rate: 2.5 times/sec. |
|
Operating Condition |
0oC
- 50oC (32oF-122oF),
0-90% R.H. (0-35oC) |
|
Power Source |
one 9V
battery |
|
Size Weight |
135mm(L)x72mm(W)x31mm(H) & 57g
Accessories: Battery & Instruction Manual |
|
* OPTIONAL TEMPERATURE PROBES |
Model
Range(Type K(CA)THERMOCOUPLE)
TP-K01 Bead -50oC to 200oC/-58oF to 392oF
General Purpose
TP-K02 Immersion -50oC to 1000oC/-58oF
to 1832oF
Liquid or gels.
TP-K03 Surface -50oC to 750oC/-58oF to
1382oF |